NeoSpectra sensors are designed to operate in a wide variety of products and environments. Sampling setups and host systems depend on the requirements of the end-use application and, hence, are developed by NeoSpectra owners.

 Core module components & typical configuration of end-usage model

Specifications’ highlights

Parameter Conditions SWS62221-1.7 SWS62221-2.1 SWS62221-2.5 Unit
Spectral Range PSD* > Max PSD/10 1,250 – 1,700 1,300 – 2,100 1,350 – 2,500 nm
Resolution @λ= 1.55 µm, FWHM criterion 16 (Optinal: 8) nm
Typical Signal to Noise Ratio  Tx** setup, 0.5 s scan time, Resolution: 16 nm 2,500 : 1 2,500 : 1 2,500 : 1
Rx*** setup, 2 s scan time, Resolution 16 nm >1,000 : 1 >1,000 : 1 >1,000 : 1
 Wavelength accuracy  @λ= 1.4 µm ; Temperature < 40oC ±1.5  nm
Wavelength repeatability @λ= 1.4 µm, absorbance level: 0.5 A.U, , Resolution: 16nm ±0.1 nm
 Temperature  Operating range -5o to 40o
(Can be configured to higher ranges upon request)
oC
table-img